KanomaxAdvanced Airflow Capture Hood

TABmaster Airflow Capture Hood Model 6720

Advanced capture hood with detachable Bluetooth micromanometer for direct volume flow plus pitot, velocity-grid and pressure measurements.

Airflow range
40–4,300 m³/h
Pressure
−2,500 to 2,500 Pa
Data memory
8,000 measurements
Wireless
Bluetooth
Detachable micromanometerBluetooth + appsAirflow / pressure / velocitySix hood configurations
CE-markHERS-approved instrumentManufacturer calibration certificate

Products

ModelGroupPositioningCore fit
6720Advanced capture hoodDirect terminal flow plus standalone pressure and velocity functionsHVAC/TAB, commissioning and critical-environment verification

Key Features

Detachable micromanometer

Use the meter on the hood or independently with pitot tubes and velocity grids.

Bluetooth workflow

Supported mobile applications enable remote control and data recording.

Broad measurement set

Measures volume flow, differential pressure, calculated velocity, temperature and humidity.

Eight-thousand-record memory

Organize extended projects and export results for reporting.

Technical Specifications

ParameterUnitValue / Description
CAPTURE HOOD
Airflow rangem³/h40–4,300
Airflow accuracy±3% of reading ±8 m³/h above 85 m³/h
Standard hoodmm610 × 610
Available hoodsSix configurations
DETACHABLE MICROMANOMETER
Pitot velocity rangem/s0.15–40
Velocity-grid rangem/s0.15–15
Differential-pressure rangePa−2,500 to 2,500
InterfacesUSB and Bluetooth
GENERAL
Data memorymeasurementsUp to 8,000
Power4 × AA batteries or AC adapter
WeightkgApprox. 3.6 with standard hood

Values shown are manufacturer specifications or documented configuration details. Confirm final model, probe, accessories, regional availability and calibration scope before ordering.

Applications

Terminal airflow balancing

Measure supply and return volumes directly at outlets.

Pitot-tube traverse

Detach the meter for in-duct velocity and flow calculation.

Pressure verification

Measure room and barrier pressure differentials.

Velocity-grid measurement

Use a grid for average face velocity at openings and filter banks.